Cone Beam Computed Tomography In Implantology 
V.N.V.Madhav 1

Address For Correspondence
Dr. V.N.V.Madhav (M.D.S)
Associate Professor.
Department of Prosthodontics,
Bharati Vidyapeeth University,
Dental college & Hospital,
Katraj-Dhankawadi,
Pune-43, Maharashtra (India)
Mobile : +91-9890251050
E-mail - vnvmadhav@yahoo.co.in 

    Abstract
Computerized tomography (CT)-based dental imaging for implant planning and surgical guidance carries both restorative information for implant positioning, as far as trajectory and distribution, and radiographic information, as far as depth and proximity to critical anatomic landmarks such as the mandibular canal, maxillary sinus, and adjacent teeth. Computed tomography imaging, also referred to as a computed axial tomography (CAT) scan, involves the use of rotating x-ray equipment, combined with a digital computer, to obtain images of the body. Using CT imaging, cross sectional images of body organs and tissues can be produced. Other imaging techniques are much more limited in the types of images they can provide. Cone Beam Computed tomography (CBCT) is a compact, faster and safer version of the regular CT. Through the use of a cone shaped X-Ray beam, the size of the scanner, radiation dosage and time needed for scanning are all dramatically reduced. A typical CBCT scanner can fit easily into any dental (or otherwise) practice and is easily accessible by patients. The time needed for a full scan is typically under one minute and the radiation dosage is up to a hundred times less than that of a regular CT scanner. In this article, the differences between the cone beam CT and conventional CT scans will be evaluated and their clinical applications in the implant therapy will be explored.

     Keywords
Computed Tomography, Cone Beam Computed Tomography, Imaging Techniques

  Full Text

 

Introduction
There has been a rapid increase in the number of practitioners involved in implant placement, including specialists and general practioners with different levels of expertise. Although the significance of accurate planning and surgical guidance as it pertains to critical anatomic landmarks such as the mandibular canal, maxillary sinus, and adjacent teeth cannot be overstated when reviewing imaging modalities for the preoperative assessment of the dental implant site, many conflicting variables need to be considered. The amount of information provided, its accuracy, and its applicability need to be weighed against cost, convenience, availability, radiation dose, and expertise required to produce and read the output of each modality. Currently there are a number of software systems that analyze computerized tomography (CT) scans to aid in planning surgery and produce the physical surgical drilling template guides. These templates are computer manufactured in such a way that they identically match the location, trajectory, and depth of the planned implant. As the dental practitioner places the implants, the templates stabilize the drilling by restricting the degrees of freedom of the drill trajectory and depth. The quantitative relationship between successful outcomes in dental implant treatment and CT-based dental imaging, coupled with surgical template guidance, is unknown and awaits discovery through large prospective clinical trials. However, using CT-based dental imaging together with surgical template guidance is becoming a reliable procedure based on a series of recent preliminary clinical studies and case reports1-8.

The development of advanced imaging in recent years is breathtaking. Just a few years ago, 3D and sectional imaging were limited to conventional helical CT. 3D reconstruction and multiplanar reformatting can only be done with CT workstations. All images have to be printed on films and viewed using light box in clinic. Today, CT scans are considered essential for multiple dental implant placements. Quite a few dental surgeons installed their own in-house CBCT largely because of high demand for dental implants. The further development of CT in dentistry will certainly be higher infiltration of CBCT machines into dental clinics and broadening its application to almost all dental treatments. Dentistry, as a whole, still needs some time to adapt to this rapid development in imaging. With the vastly improved diagnostic ability from CBCT, the treatment outcome becomes highly predictable. The quality of all dental patient care will be enhanced by it. One thing is sure: the change has just begun.

 

History
CBCT was first adapted for potential clinical use in 1982 at the Mayo Clinic Biodynamics Research Laboratory9. Initial interest focused primarily on applications in angiography in which soft-tissue resolution could be sacrificed in favor of high temporal and spatial-resolving capabilities. Since that time, several CBCT systems have been developed for use both in the interventional suite and for general applications in CT angiography.10, 11 Exploration of CBCT technologies for use in radiation therapy guidance began in 1992,12,13 followed by integration of the first CBCT imaging system into the gantry of a linear accelerator in 1999.14 

The first CBCT system became commercially available for oro-maxillofacial imaging in 2001 (NewTom QR DVT 9000; Quantitative Radiology, Verona, Italy). Comparatively low dosing requirements and a relatively compact design have also led to intense interest in surgical planning and intra-operative CBCT applications, particularly in the head and neck but also in spinal, thoracic, abdominal, and orthopedic procedures15-19.The technical and clinical considerations pertaining to CBCT imaging in many of these applications have been the subjects of several recent reviews.20-24 Commercially available CBCT systems for oro-maxillofacial imaging include the CB MercuRay and CB Throne (Hitachi Medical, Kashiwi-shi, Chiba-ken, Japan), 3D Accuitomo products (J. Morita Manufacturing, Kyoto, Japan), and iCAT (Xoran Technologies, Ann Arbor, Mich; and Imaging Sciences International, Hatfield, Pa). Similar systems designed for point-of-service head and neck imaging have also recently become available (MiniCAT, Xoran Technologies; 3D Accuitomo and 3D Accuitomo 170, J Morita Manufacturing; ILUMA Cone Beam CT, IMTEC, Ardmore, Okla and GE Healthcare, Chalfont St. Giles, UK).

 

 

Oro-Maxillofacial Imaging
Advanced cross-sectional imaging techniques such as CT are used in Oro-maxillofacial imaging to solve complex diagnostic and treatment-planning problems, such as those encountered in craniofacial fractures, endosseous dental-implant planning, and orthodontics, among others. With the advent of CBCT technology, cross-sectional imaging that had previously been outsourced to medical CT scanners has begun to take place in dental offices. 

 

Early dedicated CBCT scanners for dental use were characterized by Mozzo et al25 and Arai et al26 in the late 1990s. Since then, more commercial models have become available, inciting research in many fields of dentistry and oral and maxillofacial surgery. To date, multiple ex vivo studies have attempted to establish the ability of CBCT images to accurately reproduce the geometric dimensions of the maxillofacial structures and the mandible.27-30

A relatively low patient dose for dedicated maxillofacial scans is a potentially attractive feature of CBCT imaging. An effective dose in the broad range of 13-498 µSv can be expected, with most scans falling between 30 and 80 µSv, depending on exposure parameters. In comparison, CT with similar parameters delivers  860 µSv.31, 32 Image quality can vary considerably with dose; images acquired with higher radiation exposure often produce superior image quality. 

 

CBCT Benefits & Applications 

  • Indications of CBCT in the Maxillofacial Region
  • Evaluation of the jaw bones to assess the feasibility of placing dental implants at specific sites in the jaws. This ensures that every possible precaution has been made to reduce the risk of involvement of the nerves in the lower jaw, and the sinuses and nose in the upper jaw.
  • Evaluation of the status of previously placed implants
  • Evaluation of the hard tissue (bones) of the tempro-mandibular joint (TMJ)
  • Evaluation of abnormalities (pathology) in or affecting the bones
  • Evaluate extent of alveolar ridge resorption
  • Assessment of relevant structures prior to orthodontic treatment such as the presence and position of impacted canine and third molar teeth
  • Assessing symmetry of the face (cephalometrics)
  • Assessing the airway space (sleep apnea)
  • To permit 3D reconstructions of the bones or the fabrication of a Biomodel of the face and jaws
  • Assessing the mandibular nerve prior to the removal of impacted teeth, especially the lower wisdom teeth

 

CBCT Versus Dental X-ray

Cone beam images provide undistorted or accurate dimensional views of the jaws. Panoramic images, by contrast, are both magnified and distorted. Magnification by itself is not a problem, as long as one knows or can calculate the magnification factor. Distortion, on the other hand, is the unequal magnification of different parts of the same image. Due to distortion panoramic images are notoriously unreliable to use for making measurements33.

 

In addition, while CT images can provide cross-sectional (bucco-lingual), axial, coronal, sagittal, and panoramic views, a panoramic film provides an image of only one dimension, namely a mesio-distal or antero-posterior perspective. Further, in a panoramic image all the structures between the x-ray tube and the image detector are superimposed on one another. With CT it is possible to separate out the various structures, for example, the left condyle from the right one.

 

CBCT Compared to Tomography

Unlike panoramic radiography, plain-film tomography, if performed with the appropriate equipment, does not result in distortion. Like panoramic radiography, however, it does result in magnification, the degree of which differs from manufacturer to manufacturer. Plain-film tomography provides direct (as opposed to reconstructed) cross-sectional, sagittal and coronal views. The disadvantage of plain-film tomography is that it requires much more chair time than CT. It can thus be especially difficult to do on patients who are unable to sit or hold still for a period of time. Cone beam CT, on the other hand, can be performed within a 10-40 second range, depending on the region being imaged and on the desired quality of the image. Cone beam CT also provides stronger indication of bone quality.

 

 

CBCT Versus CT (See Table-1)

  • Cost of equipment is approximately 3-5 times less than traditional Medical CT
  • The equipment is substantially lighter and smaller.
  • Cone beam CTs have better spatial resolution (i.e. smaller pixels)
  • No special electrical requirements needed
  • No floor strengthening required
  • The room does not need to be cooled
  • Very easy to operate and to maintain; little technician training is required
  • Some cone beam manufacturers and vendors are dedicated to the dental market. This makes for a greater appreciation of the dentist 's needs
  • In the majority of cone beam CTs the patient is seated, as compared with lying down in a medical CT unit. This, together with the open design of the cone beam CTs virtually eliminates claustrophobia and greatly enhances patient comfort and acceptance. The upright position is also thought by many to provide a more realistic picture of condylar positions during a TMJ examination
  • The lower cost of the machine may be passed on to the patient in the form of lower fees
  • Both jaws can be imaged at the same time (depending on the specific cone beam machine)
  • Radiation dose is considerably less than with a medical CT.

 

Table-1 Differences between conventional CT and cone beam CTTable-1 Differences between conventional CT and cone beam CT

Table-1 Differences between conventional CT and cone beam CT

 

Imaging Modalities In Dental Implant Placement

Implantologists have long appreciated the value of 3- dimensional imaging. Conventional CT scans are used to assess the osseous dimensions, bone density, and alveolar height, especially when multiple implants are planned. Locating landmarks and anatomy such as the inferior alveolar canal, maxillary sinus, and mental foramen occurs more accurately with a CT scan. The use of the third dimension has improved the clinical success of implants and their associated prostheses, and led to more accurate and aesthetic outcomes.36-42 With CBCT technology both the cost and effective radiation dose can be reduced. CBCT has been in use in implant therapy and may be employed in orthodontics for the clinical assessment of bone graft quality following alveolar surgery in patients with cleft lip and palate.43, 44 The images produced provide more precise evaluation of the alveolus. This technology can help the clinician determine if the patient should be restored or if teeth should be moved orthodontically into the repaired alveolus. Anatomic structures such as the inferior alveolar nerve, maxillary sinus, mental foramen, and adjacent roots are easily visible using CBCT 38. The CBCT image also allows for precise measurement of distance, area, and volume. Using these features, clinicians can feel confident in the treatment planning for sinus lifts, ridge augmentations, extractions, and implant placements.

 

Before implant placement and during treatment planning, the implant clinician must be able to measure the height and width of the alveolar process to ensure adequate bone and to select appropriately sized implants. In addition, the clinician must know the precise location of the mandibular canal (injury to the neurovascular bundle within the canal can result in facial paresthesia) and the maxillary sinuses (perforation of the sinuses creates the possibility of antral infections and increases the likelihood of implant failure). Multiple views of the proposed implant site should be taken, which often require the use of different imaging procedures. Various radiographic modalities are available to the clinician, including intraoral films (i.e., periapical and occlusal radiographs), panoramic radiographs, cephalometric radiographs, plain (conventional) tomography, computed tomography (CT), cone beam CT, digital subtraction radiography, and magnetic resonance imaging.

 

Cross-sectional imaging techniques can be an invaluable tool during preoperative planning for complicated endosseous dental implantation procedures.45 Conventional linear tomography and CT have traditionally been used in presurgical imaging, though the former has overlain ghosting artifacts and the latter has relatively high radiation exposure and cost.46 

 

Practitioners have begun using office-based CBCT scanners in preoperative imaging for implant procedures, capitalizing on availability and low dosing requirements. A review by Guerrero et al47 outlines the clinical and technical aspects of CBCT, which have popularized this new technique. Preliminary evidence addresses the ability of CBCT images to characterize mandibular and alveolar bone morphology, as well as to visualize the maxillary sinuses, incisive canal, mandibular canal, and mental foramina, all structures particularly important in surgical planning for dental implantation.46,48,49 Several studies have described the 3D geometric accuracy of CBCT imaging in the maxillofacial and mandibular regions as well.50-53 

 

Limitations Of Cbct Imaging

While there has been enormous interest, current CBCT technology has limitations related to the "cone beam" projection geometry, detector sensitivity and contrast resolution. These parameters create an inherent image "noise" that reduces image clarity such that current systems are unable to record soft tissue contrast at the relatively low dosages applied for maxillofacial imaging. Another factor that impairs CBCT image quality is image artifact such as streaking, shading, rings and distortion. Streaking and shading artifacts due to high areas of attenuation (such as metallic restorations) and inherent spatial resolution may limit adequate visualization of structures in the dento-alveolar region.

 

Controversies

As with any emerging imaging technology, use of CBCT scanners has been the subject of criticism as well as acclaim55. The technology itself is limited by lack of user experience and what is currently a relatively small body of related literature. The point-of-service operational model that dominates diagnostic head and neck CBCT imaging practices has also drawn criticism. Because of the low radiation dose, CBCT can only provide bony detail and is unable to provide images of the soft tissues. Research on this technology is still preliminary, without prospective studies that convincingly demonstrate its benefit compared with conventional CT. Both in medical and oral and maxillofacial imaging in dentistry, CBCT has been largely adopted as an office-based service. This is a usage model purported to expedite patient diagnosis and treatment while simultaneously reducing costs, providing one-step management with fewer billed visits and no radiologist consultation fees. Point-of-service imaging and other self-referral services, however, have been widely criticized for encouraging overuse and directly inflating medical costs. The belief that financial incentives undermine the clinical decision-making process has been the basis for it's criticism. The advent of CBCT technologies has also fueled the controversy surrounding office-based imaging, which is usually performed and interpreted by non-radiologists often without the accreditation, training, or licensure afforded by the radiology community. 

 

Conclusions

Outcomes assessment in this area of dentistry is difficult, primarily due to bias and variability in clinical research. Observed differences can be due to differences among investigators and/or interest groups rather than differences in the treatments. Furthermore, once cost-to-benefit analyses are conducted, the increase in cost associated with CT-based implant planning and computer fabrication of surgical templates must be justified from a consumer perspective (i.e., the value associated with the increased safety and predictability of dental implants). It helps the clinician to safely and predictably transfer the optimal-implant trajectory and distances from the adjacent tooth and mandibular nerve to the patient's mouth. The final restoration becomes functional and esthetic. It does not compromise adjacent teeth or anatomic structures, yet was well accepted by the patient. CBCT is an emerging CT technology, which has potential applications for imaging of high-contrast structures in the head and neck as well as maxillofacial regions. Preliminary research suggests that high-spatial-resolution images can be obtained with comparatively low patient dose. To date, the most researched applications for head and neck CBCT are in sinus, middle and inner ear implant, and maxillofacial imaging. This technology is not without controversy, and further research is required to establish informed recommendations about its appropriate use in a clinical setting.

 

References

1. Siessegger M, Schneider BT, Mischkowski RA, et al. Use of an image-guided navigation system in dental implant surgery in anatomically complex operation sites. J Craniomaxillofac Surg. 2001;29:276-281.

2. Fortin T, Champleboux G, Bianchi S, Buatois H, Coudert JL. Precision of transfer of preoperative planning for oral implants based on cone-beam CT-scan images through a robotic drilling machine. Clin Oral Implants Res. 2002;13: 651-656.

3. Tardieu PB, Vrielinck L, Escolano E. Computer-assisted implant placement, a case report: treatment of the mandible. Int J Oral Maxillofac Implants. 2003;18: 599-604.

4. Vrielinck L, Politis C, Schepers S, Pauwels M, Naert I. Image-based planning and clinical validation of zygoma and pterygoid implant placement in patients with severe bone atrophy using customized drill guides: preliminary results from a prospective clinical followup study. Int J Oral Maxillofac Surg. 2003; 32:7-14.

5. Parel SM, Triplett RG. Interactive imaging for implant planning, placement, and prosthesis construction. J Oral Maxillofac Surg. 2004;62(suppl 2):41-47.

6. Kopp KC, Koslow AH, Abdo OS. Predictable implant placement with a diagnostic/surgical template and advanced radiographic imaging. J Prosthet Dent. 2003;89:611-615.

7. Wat PY, Chow TW, Luk HW, Comfort MB. Precision surgical template for implant placement: a new systematic approach. Clin Implant Dent Relat Res. 2002;4:88-92.

8. Van Steenberghe D, Naert I, Andersson M, Brajnovic I, Van Cleynenbreugel J, Suetens P. A custom template and definitive prosthesis allowing immediate implant loading in the maxilla: a clinical report. Int J Oral Maxillofac Implants. 2002;17:663-670.

9. Robb RA. The dynamic spatial reconstructor: an x-ray video-fluoroscopic CT scanner for dynamic volume imaging of moving organs. IEEE Trans Med Imaging 1982;1:22-33

10. Fahrig R, Fox AJ, Lownie S, et al. Use of a C-arm system to generate true three-dimensional computed rotational angiograms: preliminary in vitro and in vivo results. AJNR Am J Neuroradiol 1997;18:1507-14

11. Saint-Fe´lix D, Trousset Y, Picard C, et al. In vivo evaluation of a new system for 3D computerized angiography. Phys Med Biol 1994;39:583-95

12. SilverMD,Yahata M, Saito Y, et al. VolumeCTof anthropomorphic phantoms using a radiation therapy simulator. In: Shaw R, ed. Medical Imaging VI: Instrumentation: Proceedings of SPIE 1992;1651:197-211

13. Cho PS, Johnson RH, Griffin TW. Cone-beam CT for radiotherapy applications. Phys Med Biol 1995:40:1863-83

14. Jaffray DA, Drake DG, Moreau M, et al. A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets. Int J Radiat Oncol Biol Phys 1999;45:773-89

15. Daly MJ, Siewerdsen JH, Moseley DJ, et al. Intraoperative cone-beam CT forguidance of head and neck surgery: assessment of dose and image qualityusing a C-arm prototype. Med Phys 2006;33:3767-80

16. Rafferty MA, Siewerdsen JH, Chan Y, et al. Intraoperative cone-beam CT for guidance of temporal bone surgery. Otolaryngol Head Neck Surg 2006;134: 801-08

17. Siewerdsen JH, Moseley DJ, Burch S, et al. Volume CT with a flat-panel detector on a mobile, isocentric C-arm: pre-clinical investigation in guidance of minimally invasive surgery. Med Phys 2005;32:241-54

18. Khoury A, Siewerdsen JH, Whyne CM, et al. Intraoperative cone-beam CT for image-guided tibial plateau fracture reduction. Comput Aided Surg 2007;12: 195-207

19. Khoury A, Whyne CM, Daly M, et al. Intraoperative cone-beam CT for correction of periaxial malrotation of the femoral shaft: a surface-matching approach. Med Phys 2007;34:1380-87

20. Orth RC, Wallace MJ, Kuo MD, for the Technology Assessment Committee of the Society of Interventional Radiology. C-arm cone-beam CT: general principles and technical considerations for use in interventional radiology. J Vasc Interv Radiol 2008;19:814 -20. Epub 2008 Apr 23

21. Do¨rfler A, Struffert T, Engelhorn T, et al. Rotational flat-panel computed tomography in diagnostic and interventional neuroradiology. Rofo 2008;180: 891-98

22. Moore CJ, Am A, Marchant T, et al. Developments in and experience of kilovoltage X-ray cone beam image-guided radiotherapy. Br J Radiol 2006;79(Spec No 1):S66-78

23. Scarfe WC, Farman AG, Sukovic P. Clinical applications of cone-beam computed tomography in dental practice. J Can Dent Assoc 2006;72:75-80

24. Glick SJ. Breast CT. Annu Rev Biomed Eng 2007;9:501-26

25. Mozzo P, Procacci C, Tacconi A, et al.Anew volumetric CT machine for dentalimaging based on the cone-beam technique: preliminary results. Eur Radiol 1998;8:1558-64

26. Arai Y, Tammisalo E, Iwai K, et al. Development of a compact computed tomographic apparatus for dental use. Dentomaxillofac Radiol 1999;28:245-48

27. Lascala CA, Panella J, Marques MM. Analysis of the accuracy of linear measurements obtained by cone beam computed tomography (CBCT-NewTom). Dentomaxillofac Radiol 2004;33:291-94

28. Mischkowski RA, Pulsfort R, Ritter L, et al. Geometric accuracy of a newlydeveloped cone-beam device for maxillofacial imaging. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104:551-59. Epub 2007 Jul 5

29. Lagrave`re MO, Carey J, Toogood RW, et al. Three-dimensional accuracy ofmeasurements made with software on cone-beam computed tomography images. Am J Orthod Dentofacial Orthop 2008;134:112-16

30. Kumar V, Ludlow JB, Mol A, et al. Comparison of conventional and cone beam CT synthesized cephalograms. Dentomaxillofac Radiol 2007;36:263-69

31. Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106:106 -14. Epub 2008 May 27

32. Schulze D, Heiland M, Thurmann H, et al. Radiation exposure during midfacial imaging using 4- and 16-slice computed tomography, cone beam computed tomography systems and conventional radiography. Dentomaxillofac Radiol 2004;33:83-86

33. Serman NJ. Pitfalls of panoramic radiology in implant surgery. Ann Dent 1989;48:13-16.

34. Dula K, Mini R et al. Hypothetical mortality risk associated with spiral computed tomography of the maxilla and mandible. Eur J Oral Sci 1996; 104: 503-10

35. Brooks SL. Effective dose of two cone-beam CT scanners: i-CAT and NewTom 3G. Quarterly Publication of the American Association of Dental Maxillofacial Radiographic Technicians, Winter 2005

36. Hatcher DC, Dial C, Mayorga C. Cone beam CT for pre-surgical assessment of implant sites. J Calif Dent Assoc.2003;31:825-833.

37. Almog DM, LaMar J, LaMar FR, et al. Cone beam computerized tomography- based dental imaging for implant planning and surgical guidance, Part 1: Single implant in the mandibular molar region. J Oral Implantol. 2006;32:77-81.

38. Ganz SD. Conventional CT and cone beam CT for improved dental diagnostics and implant planning. Dent Implantol Update.2005;16:89-95.

39. Moore WS. Cone beam CT: a new tool for esthetic implant planning. Tex Dent J.2005;122:334-340.

40. Aranyarachkul P, Caruso J, Gantes B, et al. Bone density assessments of dental implant sites: 2. Quantitative cone-beam computerized tomography. Int J Oral Maxillofac Implants. 2005;20:416-424.

41. Guerrero ME, Jacobs R, Loubele M, et al. State-of-the-art on cone beam CT imaging for preoperative planning of implant placement. Clin Oral Investig.2006;10:1-7.

42. Sato S, Arai Y, Shinoda K, et al. Clinical application of a new conebeam computerized tomography system to assess multiple twodimensional images for the preoperative treatment planning of maxillary implants: case reports. Quintessence Int.2004;35:525-528.

43. Hamada Y, Kondoh T, Noguchi K, et al. Application of limited cone beam computed tomography to clinical assessment of alveolar bone grafting: a preliminary report. Cleft Palate Craniofac J. 2005;42:128-137.

44. Wortche R, Hassfeld S, Lux CJ, et al. Clinical application of cone beam digital volume tomography in children with cleft lip and palate. Dentomaxillofac Radiol. 2006;35: 88-94.

45. Tyndall DA, Brooks SL. Selection criteria for dental implant site imaging: a position paper of the American Academy of Oral and Maxillofacial Radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2000;89:630-37

46. Lofthag-Hansen S, Gro¨ndahl K, Ekestubbe A. Cone-beam CT for preoperative implant planning in the posterior mandible: visibility of anatomic landmarks. Clin Implant Dent Relat Res 2008 Sep 9. 

47. Guerrero ME, Jacobs R, Loubele M, et al. State-of-the-art on cone beam CT imaging for preoperative planning of implant placement. Clinical Oral Investig 2006;10:1-7

48. Ito K, Gomi Y, Sato S, et al. Clinical application of a new compact CT system to assess 3-D images for the preoperative treatment planning of implants in the posterior mandible: a case report. Clin Oral Implants Res 2001;12:539-42

49. Sato S, Arai Y, Shinoda K, et al. Clinical application of a new cone-beam computerized tomography system to assess multiple two-dimensional images for the preoperative treatment planning of maxillary implants: case reports. Quintessence Int 2004;35:525-28

50. Lascala CA, Panella J, Marques MM. Analysis of the accuracy of linear measurements obtained by cone beam computed tomography (CBCT-NewTom). Dentomaxillofac Radiol 2004;33:291-94

51. Mischkowski RA, Pulsfort R, Ritter L, et al. Geometric accuracy of a newly developed cone-beam device for maxillofacial imaging. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2007;104:551-59. 

52. Lagrave`re MO, Carey J, Toogood RW, et al. Three-dimensional accuracy of measurements made with software on cone-beam computed tomography images. Am J Orthod Dentofacial Orthop 2008;134:112-16

53. Kumar V, Ludlow JB, Mol A, et al. Comparison of conventional and cone beam CT synthesized cephalograms. Dentomaxillofac Radiol 2007;36:263-69

54. William C. Scarfe, Allan G. Farman, Cone beam computed tomography: A paradigm shift for clinical dentistry, Australasian Dental Practice, July/August 2007, 102-110

55. A.C. Miracle, S.K. Mukherji, Conebeam CT of the Head and Neck, Part 2:Clinical Applications, Aug 2009Am J Neuroradiol 30:1285-92  


  « Previous Article ToC Next Article »